Cryogenic Concentration System
The Background
Groundwater keeps a record of when it last touched the atmosphere, written in trace gases like sulfur hexafluoride (SF6) and chlorofluorocarbons (CFCs). Reading that record means measuring concentrations in the femtomole per liter range, far below what any detector can see in a raw water sample.
The Problem
For years, a geoscience research organization did this on a bench-top system assembled in house and run by hand, with every sparge, trap, and transfer step depending on the person at the bench. They asked Wasson-ECE to turn that method into a less hands-on instrument.
The Wasson-ECE Solution
Our answer was a fully automated cryogenic concentration front end that strips the dissolved gases from up to half a liter of water, dries and traps them at -75 °C using liquid nitrogen, and delivers them to the column as a single sharp injection, with the whole sequence from sample load to final report running unattended under our Hermes software. The result is a system a new analyst can operate on day one, with documented methods, replaceable components, and factory support behind it. In verification, it resolved SF6 and Halon 1301 at 1.0 fmol/L and CFC-113 at 0.02 pmol/L with retention times repeatable to better than 0.1% RSD across eight runs. Concentration problems like this one rarely have an off-the-shelf solution. If your matrix is hiding the analytes you need, we would like to hear about it.