Biogas Analysis

Analyzer Description

Configuration:

FID, FID, TCD, TCD, 8-valves, 6-packed columns, 4-capillary columns, deans switch

Sample Type:

Biogas

Compounds Separated:

C1 through C5 paraffins and olefins

C5 through C20 paraffins

Benzene, toluene, ethylbenzene, m/p-xylene composite, o-xylene, phenol

Carbon dioxide, argon/oxygen composite, nitrogen, methane, carbon monoxide, hydrogen, helium

Lower Detection Limits:

10 ppm for hydrocarbons

80 ppm for backflush

100 ppm for hydrogen, helium, argon/oxygen composite

200 ppm for nitrogen, carbon dioxide

400 ppm for carbon monoxide

Methods Met:

Wasson-ECE W3208

Typical Chromatogram

FID A ChromatogramFID B ChromatogramTCD Chromatogram

Key Features and Benefits

  • Can be configured with an additional GC by heated umbilical for simultaneous analysis of hydrogen sulfide, carbonyl sulfide, ammonia cyanide, and hydrogen cyanide.
  • Cryogenic cooling may be used to separate argon and oxygen.
  • The backflush peak can be tailored to specific samples (e.g. C6+ or C8+ backflush composite peaks).
  • Separate sampling loops eliminate cross contamination.
  • Can be configured with an injector oven for on-line analysis.
  • Fast analysis with hydrocarbon analysis through 1,3-butadiene is complete in under 10 minutes.
  • Detailed analysis with hydrocarbon analysis through n-eicosane is complete in approximately 30 minutes.