Hydrocarbon and Oxygenate Analysis in Water or Butanes

Analyzer Description

Configuration:

FID, FID, 1-valve, 4-capillary columns, deans switch

Sample type:

Mixed butane/butylenes

Butanes

Water

Combination of sample types by static mixer

Compounds Separated:

  • methane
  • ethane
  • ethylene
  • propane
  • propylene
  • isobutane
  • butane
  • trans-2-butene
  • 1-butene
  • isobutylene
  • cis-2-butene
  • isopentane
  • n-pentane
  • 1,3-butandiene
  • Isopropanol
  • tert-butanol
  • 2-butanol
  • 1-butanol

Lower detection limits:

20 ppm

Methods met:

Wasson-ECE W3583

Typical Chromatogram

FID A ChromatogramFID B Chromatogram

Key Features and Benefits

  • Can be configured as an online process instrument for analysis of multiple streams.
  • Online process instrument is constructed for safe automated operation in a Class 1, Division 2, Groups C & D areas.
  • Analyzer can be configured as a stand alone unit using liquid sampling valves or auto-injection for liquid sample introduction.
Corporate Headquarters

Wasson-ECE Instrumentation

101 Rome Ct., Fort Collins, Colorado USA 80524

Phone: +1 (970) 221-9179

Fax: +1 (970) 221-9364

© 2026. All Rights Reserved.