Nitrogen Containing Compounds in Ethylene, Propylene, Butane, and Pentane Streams

Analyzer Description

Configuration:

3-valves, 2-capillary columns, on-board vaporizer, NCD

Sample type:

Ethylene, propylene, butane, pentane

Compounds Separated:

Methylamine, dimethylamine, trimethylamine, ethylamine, diethylamine, propylamine, pyridine, aniline, quinoline, indole

Lower detection limit:

0.5 ppm

Methods met:

Wasson-ECE W4043

Typical Chromatogram

NCD ChromatogramNCD ChromatogramNCD Chromatogram

Key Features and Benefits

  • Configured for both liquid and gas sample analysis.
  • On-board vaporizer converts liquid to gas phase prior to injection for more repeatable analysis.
  • High selectivity and equimolar for nitrogen.
  • A 3-way decoking valve alleviates any coking problems that might occur in the NCD burner during the oxidation process and ensures maximum response and consistency.