Polymer Grade Ethylene and Propylene Impurity Analysis by FID
Analyzer Description
Configuration:
5-valve, 3-column (capillary), internal vaporizer, methanizer, FID / FID
Sample Type:
Polymer grade ethylene and propylene
Components Separated:
- carbon monoxide
- carbon dioxide
- methane
- ethane
- ethylene
- propane
- propylene
- isobutene
- acetylene
- n-butane
- propadiene
- trans-2-butene
- 1-butene
- isobutene
- cis-2-butene
- isopentane
- methyl acetylene
- n-pentane
- 1,3-butadiene
Lower Detection Limit:
Carbon monoxide, carbon dioxide, methane: 0.05 ppm
1 ppm for all other components
Methods Met:
Wasson-ECE 262, Wasson-ECE 260 (ethylene only), Wasson-ECE 261 (propylene only)
Typical Chromatogram



Key Features and Benefits
- Separate sampling loops eliminate cross contamination
- On-board vaporizer converts propylene to gas phase prior to injection for more repeatable analyses