Pyrolysis Gas Analysis

Analyzer Description

Configuration:

FID, FID, 5-valves, 2-capillary columns

Sample type:

Pyrolysis gas

Compounds Separated:

C1 - C6 hydrocarbons

C7 backflush

Benzene, toluene, styrene, decane, butylated hydroxytoluene (BHT)

Lower detection limits:

10 ppm for butylated hydroxytoluene (BHT)

1 ppm for remaining compounds

Methods met:

Wasson-ECE W3804

Typical Chromatogram

FID A ChromatogramFID B Chromatogram

Key Features and Benefits

  • Can be configured for both liquid and gas sample analysis.
  • The backflush peak can be tailored to specific samples (e.g. C6+ or C8+ backflush composite peaks).